PAT700 TOC Dual Stream

Product No:PAT700220

The PAT700 TOC analyzer fully complies with all global pharmacopeias (USP, EP, JP) requirements. Each sample is fully oxidized guaranteeing the most accurate TOC result that follows EP2.2.44 guidelines. Production uptime is assured with the PAT700 main and standby dual UV lamp design, making it the ideal choice for on-line TOC release. Save time and money - the PAT700 can be installed online and run up to 4 grab samples offline.

规格

Platform Anatel
Industry Standards 21 CFR Part 11 EP <2.2.44> USP <643>
Calibration Intervals and Options 3-point TOC calibration, Conductivity calibration to USP<645>, System suitability to USP and EP
Fluid Compatibility Purified water, Water for injection (WFI)
Online Operation Sample inlet pressure 10 to 100psi (0.69 to 6.89Bar)
Interfaces Protocols 3 x 4-20mA outputs │ 2 isolated digital inputs │ 4 x isolated digital outputs , Ethernet
Detection Differential conductivity
IP Rating IP46
Reporting ppb TOC │ µS/cm
Installation Category II
功率要求 100/230VAC ± 10%, 50/60Hz, 65W
分辨率 0.1 ppb
Operating Modes Online TOC, Online Conductivity, Auto-switching Main and Standby UV lamps for 12 months operation │ 4-bottle auto-sampler for grab sample analysis │ Excursion capture for root cause analysis support
重量 13.6 kg
宽度 59.7 cm
深度 22.9 cm
高度 25.4 cm
Color Blue
Package Quantity 1