We and our partners use cookies and other tracking technologies to collect data relating to you to perform analytics, improve your experience of using our website, provide you with personalized ads and content based on your interactions with these and other websites and allow you to share content on social media. By clicking “Accept All Cookies”, you consent to this and to the sharing of this data with our partners. You can change your consent preferences at any time in the “Cookie Settings” section at the bottom of our website. Review our Cookie Notice to learn more about our practices.
What is required to comply with the USP chapter on Conductivity (USP <645>) for a TOC analyzer?

For a TOC analyzer to report conductivity in compliance with USP<645>, it must be possible to calibrate both the cell constant and the meter accuracy. The PAT700 is designed to comply fully with the requirements of USP <645>.
In USP <645> for Conductivity, the calibration requirements include checking the conductivity instrument’s cell constant using a traceable, certified conductivity solution, as well as checking the meter accuracy using a NIST-certified resistor. The PAT700 allows the use of a traceable, certified conductivity solution to check its cell constant and also the use of a NIST-certified resistor to check the accuracy of its meter. Hence it is fully compliant and can be used to report both TOC to USP <643> and Conductivity to USP <645>.