We and our partners use cookies and other tracking technologies to collect data relating to you to perform analytics, improve your experience of using our website, provide you with personalized ads and content based on your interactions with these and other websites and allow you to share content on social media. By clicking “Accept All Cookies”, you consent to this and to the sharing of this data with our partners. You can change your consent preferences at any time in the “Cookie Settings” section at the bottom of our website. Review our Cookie Notice to learn more about our practices.
SterilGARD for CytoFLEX SRT Biosafety Cabinet
Class II, Type Biosafety Containment Cabinet designed specifically for the CytoFLEX SRT Cell Sorter through a partnership with Baker Corporation. The biosafety cabinet relies on the flow of air to provide the following protection:
- Personnel protection or containment by an intake air velocity of no less than 100 feet per minute (fpm) through the front access opening.
- Product protection by the HEPA (high efficiency particulate air) filtered downflow air inside the cabinet work area.
- Environmental protection by allowing only HEPA filtered to be exhausted to the room.
All microbiological tests were performed using test methods and acceptance guidelines set forth by the following National and International Biological Safety Cabinet Standards which are to be following all regulatory biosafety requirements internationally.
- NSF/ANSI International Standard 49- 2019
- European Standard (EN 12469:2000)
- British Standard (BS EN 12469:2000)
- South Africa National Standard (SANS 12469:2001)
- French Standard (NF-095:2006)
- China Standard (SFDA YY- 0569:2011)
- Japanese Industrial Standard (JIS K 3800:2009)
- Australian Standard (AS 1807.1:2009)